NEW & REFURBISHED MEDICAL EQUIPMENT sale 240910
• FRONT FACING : a screen of parameters, selectors and indicators.
• RIGHT SIDE FACADE : pneumatic connections, electrical connection.
• THE REAR FACADE : ambient air intake, NiCd battery access hatch.
VERSATILE RESPIRATOR.
• Reduced dimensions (250 x 210 x 170 mm)
• Weight of 5 kg with carrying bag.
The shell is made of “shock” ABS with integrated carrying handle.
VENTILATION MODES:
Controlled ventilation (VAC), assisted volume-controlled ventilation (ACMV), spontaneous pressure-assisted ventilation (VS-AI/PS).
VAC ACMV
FiO2 adjustable either in pure oxygen or in an oxygen/air mixture (60%). Tidal volume (VT) adjustable from 100 to 1500 ml. Frequency adjustable from 6 to 40 c/min. The I/E ratio adjustable from 1/1 to 1/3. The integrated Positive Expiratory Pressure (PEEP) is adjustable from 0 to 10 cm H2O.
In ACMV mode, a pre-set tidal volume is blown into the patient circuit with each inspiratory effort, using a constant flow rate. In the event of apnea, the device will operate on preset safety settings.
VS-AI/PS
Inspiratory support (constant positive pressure), adjustable from 5 to 40 cm H20, is obtained with each inspiratory effort. Positive Expiratory Pressure is generally set between 7 and 10 cm H2O.
OPERATING PRINCIPLES:
The oxygen supply pressure is 3.5 bars (+/- 0.7) and the consumption specific to the device is approximately 1l/min. The internal batteries, essential for powering the electronic circuit, have a battery life of 10 hours and are recharged in 10 to 14 hours. Osiris 2 can be plugged into the mains.