NEW & REFURBISHED MEDICAL EQUIPMENT sale 240718

WEINMANN MEDUVENT STANDARD MOBILE RESPIRATOR on LIFE-BASE XS transport station EMERGENCY VENTILATOR (INVASIVE AND NON-INVASIVE) NEW (made in GERMANY) NEUF

L'asta comincerà tra __ giorni e __ ore

Prezzo base: €1,000

Prezzo stimato: €3 000 - €5 000

IVA: Solo su commissione

Product class according to directive 93/42/EEC: IIb

Dimensions (WxHxD): 206mm x 138mm x 130mm

Weight :

Without battery: Approx. 2kg

With battery: Approx. 2.5kg

Weight with Capnography option:

Without battery: Approx. 2.15kg

With battery; Approx. 2.65kg

Functioning :

Temperature range: -20°C to +50°C

Air humidity: 0% RH to 95% RH, non-condensing

Atmospheric pressure: 540 hPa to 1100 hPa

Altitude: from -500 m to 5,000 m

Storage (device)/Transport:

Temperature range: -40°C to +70°C

Air humidity: 0% RH to 95% RH, non-condensing

Atmospheric pressure: 540 hPa to 1100 hPa

Altitude: from -500 m to 5,000 m

Power supply (nominal voltage): 12 V

Permissible operating voltage 10.2 V to 15.1 V

Maximum power consumption 30 W

Mains disconnection Remove the mains plug to completely switch off the device

of the sector.

Current consumption 0.1 to 3 A

Input voltage (power supply/charger): 100 V/240 V~-50 Hz/60 Hz

Radio waves (only with Bluetooth® data transfer option)

Frequency band: 2.4 GHz to 2.57 GHz

Transmitting power: 12 dBm max.

Modulation modes:

• 1 Mb/s: GFSK (BDR)

• 2 Mb/s: π/4-DQPSK (EDR)

• 3 Mb/s: 8-DPSK (EDR)

Shock and vibration resistance

• IN 1789

• EN 60601-1-12 (categories: fixing in a vehicle

rescue, fixing in an airplane, fixing in a helicopter,

at different locations at the emergency scene)

• EUROCAE ED-14G (RTCA DO 160 G) (Section 8 – Vibration

Cat aircraft. S and Cat helicopters. U2)

• MIL-STD 810 G (Cat. 12, 13, 14, 20)

Type of emergency vehicle

Attachment in emergency vehicle, boat, airplane and

helicopter, as well as at different locations on the scene of the emergency

Display screen

5" TFT color screen

Resolution 320 pixels x 240 pixels

Alarm volume

from 60 dBA to 95 dBA for all alarm priorities and all alarms

Standards applied

EN 60601-1; -1-2; -1-6; -1-8; -1-12

IN 1789

EN 794-3

ISO 10651-3

RTCA DO-160 G

MIL-STD 810 G

EN ISO 80601-2-55 (only with Capnography option)

With the Transfer option. Bluetooth® data:

EN 62311

EN 301489-1

EN 300328

Volume-controlled ventilation modes

VC, CPR, ISR

Optional: SIMV (only with SIMV option), SIMV/AI

(only with the SIMV option and the Flow measurement/AI option),

VAC (only with VAC option)

Pressure-controlled ventilation modes

VS-PEP, Request

Optional: VS-PEP/AI (only with Flow measurement/AI option)

PCV (only with Pressure Controlled Ventilatory Modes option)

aVPC (only with Pressure Controlled Ventilatory Modes option)

BiLevel/AI (only with Pressure Controlled Ventilatory Modes option)

VCRP/AI (only with Pressure Controlled Ventilatory Modes option)

CCSV (only with CCSV option)

Inhalation flow (only with the Inhalation option) from 0 l/min to 10 l/min, in steps of 1 l/min

Monitoring

Pressure gauge:

Airway pressure

Frequency meter (only with CCSV option)

Curves:

• Airway pressure (only with option

Flow measurement/AI and curve display option or option

Capnography)

• Flow (only with the Flow measurement/AI option and the Curve display option)

• CO2 (only with the Capnography option)

• Tend. etCO2 (only with Capnography option)

Monitoring Settings

pPeak, pFlat, pAvg

Vte, VMe, f, fsp, Vleak (only with the Measurement of

flow/AI)

etCO2 (only with Capnography option)

Gas used

Medical oxygen (100% oxygen)

Or

Concentrator oxygen compliant with pharmacopoeia

European (90% to 96% oxygen)

Operating pressure from 2.7 bars to 6 bars

Recommended gas supply 4.5 bar for ≥ 150 l/min

Required gas supply at least 2.7 bar for ≥ 80 l/min

Maximum output flow

80 l/min for a supply pressure of 4.5 bars in mode

Air Mix and No Air Mix mode

Mechanical emergency air/pressure relief valve

Pressure limit at 100 mbar maximum

IE

1/1 in RCP 30/2, 15/2 and manual ISR from 1/4 to 4/1*

* Available I/E settings in ventilation modes

in controlled volume depend on the frequency combination

ventilation/tidal volume.

Ventilatory rate from 5 min-1 to 50 min-1 (±1 min-1)

Inspiratory time from 0.2 s to 9.6 s

Tidal volume(1) from 50 ml to 2000 ml (±40 ml or ±20%)

Breathing volume per minute(1) 0.25 l minimum

20 liters maximum

Gas composition

Mixture of air, oxygen and CO2. Percentage of oxygen between

21% and 100%, CO2 percentage between 0% and 10%

Inspiratory pressure (pInsp) (only with Pressure Controlled Ventilatory Modes option)

from 3 mbar to 60 mbar (±3 mbar or ±15%)

Maximum ventilation pressure (pMax)

from 10 mbar to 65 mbar (±3 mbar or ±15%)

Minimum operating pressure 3 mbar. Impossible to adjust.

Minimum limit pressure (pLim, min) 10 mbar. The device does not produce active depression.

Maximum limit pressure (pLim, max) 65 mbar

Pressure limit method Pressure regulation

Method guaranteeing the minimum value

Pressure regulation

PEEP from 0 mbar to 30 mbar (±3 mbar or ±15%)

PEP CCSV

(only with CCSV option) from 0 mbar to 5 mbar (±3 mbar or ±15%)

Inspiratory support ΔpAI (only

with Flow measurement/AI option)

from 0 mbar to 35 mbar (±3 mbar or ±15%) via PEP

Pressure ramp (only with Flow measurement/AI option)

Steep

Average

Gentle

Trigger (fixed setting)(1)

• Inspiratory trigger:

-1.2 mbar for PEEP > 0

-0.8 mbar for PEEP = 0

• Expiratory trigger:

30% of maximum flow

Trigger via internal sensors

Adjustable trigger (only with Flow measurement/AI option)(1)

Trigger via FlowCheck sensor

Inspiratory trigger (3 levels):

• Level 1: sensitive, corresponds to approx. 3 l/min.

• Level 2: medium sensitivity, corresponds to approximately 5 l/min.

• Level 3: insensitive, corresponds to approx. 10 l/min.

Expiratory trigger (3 levels):

• Level 1: long AI inflation, corresponds to approx. 10%. of maximum inspiratory flow

• Level 2: average AI inflation, corresponds to approximately 35% of the maximum inspiratory flow

• Level 3: short AI inflation, corresponds to approx. 70%. of maximum inspiratory flow

Inspiratory trigger (units): from 1 l/min to 15 l/min

Expiratory trigger (units): from 5% to 80% of maximum flow.

Adjustable trigger (only with CCSV option)

Inspiratory trigger (5 levels):

• Level 1: very sensitive (low pressure must be exerted on the thorax to trigger inflation) up to

• Level 5: non-sensitive (firm pressure must be exerted on the thorax to trigger inflation)

Trigger window for controlled respiratory movements

20% Te (SIV, SIMV/AI, BiLevel/AI and VCRP/AI)

0% to 100% Te (VAC)

AI respiratory movement trigger window

from 0% to 100% Te

Fen. aVPC trigger (only with the Pressure-controlled ventilation modes option)

from 0% to 100% Te (adjustable)

Volume monitoring(1 ) Measuring range: 40 ml to 8,000 ml

Tolerance: ±15%

Airway pressure monitoring

Measuring range: -5 mbar to +80 mbar

Tolerance: ±3 mbar

Oxygen concentration:

• Air Mix mode

• No Air Mix mode: 100% oxygen

Concentrator oxygen (90% to 96% oxygen)

Thread for gas under pressure External thread G 3/8