NEW & REFURBISHED MEDICAL EQUIPMENT sale 240718
Product class according to directive 93/42/EEC: IIb
Dimensions (WxHxD): 206mm x 138mm x 130mm
Weight :
Without battery: Approx. 2kg
With battery: Approx. 2.5kg
Weight with Capnography option:
Without battery: Approx. 2.15kg
With battery; Approx. 2.65kg
Functioning :
Temperature range: -20°C to +50°C
Air humidity: 0% RH to 95% RH, non-condensing
Atmospheric pressure: 540 hPa to 1100 hPa
Altitude: from -500 m to 5,000 m
Storage (device)/Transport:
Temperature range: -40°C to +70°C
Air humidity: 0% RH to 95% RH, non-condensing
Atmospheric pressure: 540 hPa to 1100 hPa
Altitude: from -500 m to 5,000 m
Power supply (nominal voltage): 12 V
Permissible operating voltage 10.2 V to 15.1 V
Maximum power consumption 30 W
Mains disconnection Remove the mains plug to completely switch off the device
of the sector.
Current consumption 0.1 to 3 A
Input voltage (power supply/charger): 100 V/240 V~-50 Hz/60 Hz
Radio waves (only with Bluetooth® data transfer option)
Frequency band: 2.4 GHz to 2.57 GHz
Transmitting power: 12 dBm max.
Modulation modes:
• 1 Mb/s: GFSK (BDR)
• 2 Mb/s: π/4-DQPSK (EDR)
• 3 Mb/s: 8-DPSK (EDR)
Shock and vibration resistance
• IN 1789
• EN 60601-1-12 (categories: fixing in a vehicle
rescue, fixing in an airplane, fixing in a helicopter,
at different locations at the emergency scene)
• EUROCAE ED-14G (RTCA DO 160 G) (Section 8 – Vibration
Cat aircraft. S and Cat helicopters. U2)
• MIL-STD 810 G (Cat. 12, 13, 14, 20)
Type of emergency vehicle
Attachment in emergency vehicle, boat, airplane and
helicopter, as well as at different locations on the scene of the emergency
Display screen
5" TFT color screen
Resolution 320 pixels x 240 pixels
Alarm volume
from 60 dBA to 95 dBA for all alarm priorities and all alarms
Standards applied
EN 60601-1; -1-2; -1-6; -1-8; -1-12
IN 1789
EN 794-3
ISO 10651-3
RTCA DO-160 G
MIL-STD 810 G
EN ISO 80601-2-55 (only with Capnography option)
With the Transfer option. Bluetooth® data:
EN 62311
EN 301489-1
EN 300328
Volume-controlled ventilation modes
VC, CPR, ISR
Optional: SIMV (only with SIMV option), SIMV/AI
(only with the SIMV option and the Flow measurement/AI option),
VAC (only with VAC option)
Pressure-controlled ventilation modes
VS-PEP, Request
Optional: VS-PEP/AI (only with Flow measurement/AI option)
PCV (only with Pressure Controlled Ventilatory Modes option)
aVPC (only with Pressure Controlled Ventilatory Modes option)
BiLevel/AI (only with Pressure Controlled Ventilatory Modes option)
VCRP/AI (only with Pressure Controlled Ventilatory Modes option)
CCSV (only with CCSV option)
Inhalation flow (only with the Inhalation option) from 0 l/min to 10 l/min, in steps of 1 l/min
Monitoring
Pressure gauge:
Airway pressure
Frequency meter (only with CCSV option)
Curves:
• Airway pressure (only with option
Flow measurement/AI and curve display option or option
Capnography)
• Flow (only with the Flow measurement/AI option and the Curve display option)
• CO2 (only with the Capnography option)
• Tend. etCO2 (only with Capnography option)
Monitoring Settings
pPeak, pFlat, pAvg
Vte, VMe, f, fsp, Vleak (only with the Measurement of
flow/AI)
etCO2 (only with Capnography option)
Gas used
Medical oxygen (100% oxygen)
Or
Concentrator oxygen compliant with pharmacopoeia
European (90% to 96% oxygen)
Operating pressure from 2.7 bars to 6 bars
Recommended gas supply 4.5 bar for ≥ 150 l/min
Required gas supply at least 2.7 bar for ≥ 80 l/min
Maximum output flow
80 l/min for a supply pressure of 4.5 bars in mode
Air Mix and No Air Mix mode
Mechanical emergency air/pressure relief valve
Pressure limit at 100 mbar maximum
IE
1/1 in RCP 30/2, 15/2 and manual ISR from 1/4 to 4/1*
* Available I/E settings in ventilation modes
in controlled volume depend on the frequency combination
ventilation/tidal volume.
Ventilatory rate from 5 min-1 to 50 min-1 (±1 min-1)
Inspiratory time from 0.2 s to 9.6 s
Tidal volume(1) from 50 ml to 2000 ml (±40 ml or ±20%)
Breathing volume per minute(1) 0.25 l minimum
20 liters maximum
Gas composition
Mixture of air, oxygen and CO2. Percentage of oxygen between
21% and 100%, CO2 percentage between 0% and 10%
Inspiratory pressure (pInsp) (only with Pressure Controlled Ventilatory Modes option)
from 3 mbar to 60 mbar (±3 mbar or ±15%)
Maximum ventilation pressure (pMax)
from 10 mbar to 65 mbar (±3 mbar or ±15%)
Minimum operating pressure 3 mbar. Impossible to adjust.
Minimum limit pressure (pLim, min) 10 mbar. The device does not produce active depression.
Maximum limit pressure (pLim, max) 65 mbar
Pressure limit method Pressure regulation
Method guaranteeing the minimum value
Pressure regulation
PEEP from 0 mbar to 30 mbar (±3 mbar or ±15%)
PEP CCSV
(only with CCSV option) from 0 mbar to 5 mbar (±3 mbar or ±15%)
Inspiratory support ΔpAI (only
with Flow measurement/AI option)
from 0 mbar to 35 mbar (±3 mbar or ±15%) via PEP
Pressure ramp (only with Flow measurement/AI option)
Steep
Average
Gentle
Trigger (fixed setting)(1)
• Inspiratory trigger:
-1.2 mbar for PEEP > 0
-0.8 mbar for PEEP = 0
• Expiratory trigger:
30% of maximum flow
Trigger via internal sensors
Adjustable trigger (only with Flow measurement/AI option)(1)
Trigger via FlowCheck sensor
Inspiratory trigger (3 levels):
• Level 1: sensitive, corresponds to approx. 3 l/min.
• Level 2: medium sensitivity, corresponds to approximately 5 l/min.
• Level 3: insensitive, corresponds to approx. 10 l/min.
Expiratory trigger (3 levels):
• Level 1: long AI inflation, corresponds to approx. 10%. of maximum inspiratory flow
• Level 2: average AI inflation, corresponds to approximately 35% of the maximum inspiratory flow
• Level 3: short AI inflation, corresponds to approx. 70%. of maximum inspiratory flow
Inspiratory trigger (units): from 1 l/min to 15 l/min
Expiratory trigger (units): from 5% to 80% of maximum flow.
Adjustable trigger (only with CCSV option)
Inspiratory trigger (5 levels):
• Level 1: very sensitive (low pressure must be exerted on the thorax to trigger inflation) up to
• Level 5: non-sensitive (firm pressure must be exerted on the thorax to trigger inflation)
Trigger window for controlled respiratory movements
20% Te (SIV, SIMV/AI, BiLevel/AI and VCRP/AI)
0% to 100% Te (VAC)
AI respiratory movement trigger window
from 0% to 100% Te
Fen. aVPC trigger (only with the Pressure-controlled ventilation modes option)
from 0% to 100% Te (adjustable)
Volume monitoring(1 ) Measuring range: 40 ml to 8,000 ml
Tolerance: ±15%
Airway pressure monitoring
Measuring range: -5 mbar to +80 mbar
Tolerance: ±3 mbar
Oxygen concentration:
• Air Mix mode
• No Air Mix mode: 100% oxygen
Concentrator oxygen (90% to 96% oxygen)
Thread for gas under pressure External thread G 3/8